 |
 |

Morbidity Associated With Prolonged Use of Polyvinyl Feeding Tubes
Edward G. Hayhurst, MD;
Michael Wyman, MD
Am J Dis Child. 1975;129(1):72-74.
Abstract
In two newborn infants, morbidity was associated with long-term use of polyvinyl chloride feeding tubes. In one case, a hard, sharp catheter was thought to be responsible for peritonitis. In the other, the polyvinyl tube hardened into a coil that could only be extricated by introduction of an intraluminal guide wire and prolonged manipulation under fluoroscopic control. The rigidity acquired by these tubes is thought to present serious potential for morbidity and mortality. Methods and materials appropriate for short-term alimentation should not be presumed appropriate for long-term use, and alternatives should be considered.
Author Affiliations
From the Department of Pediatrics and Communicable Diseases, University of Michigan Medical Center, Ann Arbor. Dr. Hayhurst is now with the University of Arizona Medical Center, Tucson.
Footnotes
Received for publication May 24, 1974; accepted June 19.
Reprint requests to Department of Pediatrics and Communicable Diseases, University of Michigan Medical Center, 1405 E Ann St, Ann Arbor, MI 48104 (Dr. Wyman).
CiteULike Connotea Del.icio.us Digg Reddit Technorati Twitter
What's this?
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
Subjective Distress during Continuous Enteral Alimentation: Superiority of Silicone Rubber to Polyurethane
Herrmann et al.
JPEN J Parenter Enteral Nutr 1989;13:281-285.
ABSTRACT
Complications Occurring during Enteral Nutrition Support: A Prospective Study
Cataldi-Betcher et al.
JPEN J Parenter Enteral Nutr 1983;7:546-552.
ABSTRACT
Sustained Flexibility in Infant Feeding Tubes Containing Nonmigrating Plasticizer
Pereira et al.
JPEN J Parenter Enteral Nutr 1982;6:64-67.
ABSTRACT
Polyurethane Tubing for Transpyloric Intubation
SAUNDERS et al.
Arch Pediatr Adolesc Med 1977;131:811-812.
ABSTRACT
|